In examining the behaviour of both the limb radiances and
ozone ACDs, it appears as if the formula for calculating the limb
elevation angle given by equation (3.1) may not be
sufficiently accurate. This necessitates calibration of the
elevation angle using the limb radiances.
Starting with equation (3.1) and adding a term for refraction,
as discussed in Section 4.7, as well as a constant
offset correction term, the absolute elevation angle
is11.1,
The elevation angle calibration process makes use of forward model
calculations. Essentially, measured radiances are shifted
until they are aligned with the model radiances and the shift in EA
at which this occurs is
.
The limb radiances used in the calibration are taken from
21 September 1997, the Fairbanks to Hawaii transit flight.
Three successive scans are used from 2248 UTC
(35.8
N, 155.0
W) to 2355 UTC
(28.2
N, 155.5
W) at a 5 nm integrated
band centered at 747.5 nm. Three scans are used to reduce the
impact of random errors associated with the measurements.
These scans were chosen as both the
CPFM-derived and TOMS-derived albedo were less than 0.2 throughout
the scans which suggests little or no cloud cover.
In addition, the SZA angle varied by only 1.4
and the ER-2
altitude by 0.4 km over the course of the scans.
For purposes of calibration, the model elevation
angles will be adjusted as opposed to those of the measurements
(which will be the normal procedure).
The results of the calibration are shown in
Figure 6.26. Radiances from the individual scans
are consistent with each other indicating that atmospheric
conditions have changed very little throughout these scans.
The shift required to align the measurements was
.
Also shown in Figure 6.26
are the model radiances at the unrefracted elevation angles, although
it was hard to differentiate as the largest refraction correction
was about 0.2
.
In practice, the calibration was done
by first interpolating the model radiances (calculated at 1
increments) to the elevation angles of the measurements,
shifted by a specified
amount. The RMS-difference was computed and the amount of
the shift was varied in 0.1
increments until a minimum was
reached. Several values of total aerosol optical depth were
used as this will impact both the magnitude and the curvature of the
limb radiances. Also varied was the surface albedo. An aerosol
optical depth of 0.015 and an albedo of 0.1 were found to give the
best agreement.
To assess how sensitive the calibration is to aerosol optical depth,
the limb-viewing angle was calibrated for optical depths of
0.01 and 0.02. The results are shown in Figure 6.27.
To calibrate to the 0.02 optical depth case, an additional
shift of 0.5
was required while calibration to the 0.01
optical depth case required a 0.5
shift decrease,
relative to that for the optimal 0.015 case. However, by
observing the dashed lines in Figure 6.27, neither
is able to capture the curvature of the measured radiances.
This is the reason that the RMS-difference
was a minimum for the 0.015 optical depth case.
As the difference in
is large (0.5
between
aerosol optical depths of 0.01 and 0.015 and between 0.015 and 0.02;
recall refraction corrections were no larger than 0.2
),
the correct aerosol profile must be used to
properly calibrate the limb-viewing angle.
One important question which needs to be addressed in future
work is can two markedly different aerosol profiles and hence
different values of
give comparable
minimum least-squares differences?
If the only aerosol
profile variable in question were the total optical depth, the answer would
be no. However, shape of the profile and altitude of the peak,
for example, are variables that have not been examined but will also
impact the calibration.
From Chapter 5, it appears that three or four pieces of profile
information are available from the limb radiances. It should be possible
to substitute one of these pieces for
.
That is, retrieve a, b, c, and
together
(where a, b, and c are fitting coefficients as discussed
in Section 5.4).